Particle Aerodynamic
Size Separator (PASS)
Introduction
Measurement
of aerosol particle size distribution is important in a host
of fields such as inhalation hazard evaluation, atmospheric
studies, characterizing lung delivery systems, material synthesis
etc. Cascade impactors are the most rugged and widely used
systems all over the world for this purpose. However, most
commercially available impactors are expensive and in order
to make them available at a cost-effective scale, the Particle
Aerodynamic Size Separator (PASS), has been designed and developed.
It is an import substitute product for aerodynamic particle
sizing, based on the principle of inertial impaction and separates
particles according to their aerodynamic diameters in the
range of 0.53-10 µm in seven class intervals. Being
indigenous in design, it may be fabricated on a commercial
scale at a cost much less than that of the imported units.
This instrument will be of help to several national laboratories
and universities for carrying out studies on atmospheric pollution
and aerosol characterization.
Specification
- Flow rate
- Maximum number of stages
- Effective Cut-off diameter range
|
:
:
:
|
45
lpm
seven + Back-up filter
0.53 to 10 µm |
Applications
- This instrument is useful for
occupational monitoring of radioactive and non radioactive
environments, arising in the context of material processing,
metal cutting, powder handling etc.
- It is useful for air quality regulators
as well as for researchers engaged in various aspects of
air pollution and aerosol research.
- It can also be used by the pharmaceutical
industries for characterizing the lung aerosol delivery
systems, such as nebulizers and metered dose inhalers
Contact for
more information
Head,
Technology Transfer & Collaboration Division,
Bhabha Atomic Research Centre,
Trombay, Mumbai - 400 085
Tel : 091-022-25505337/25593897
Fax : 091-022-25505151
E-mail : headttcd@barc.gov.in

|