Cardiac Output Monitor reveals pump
performance of the heart. The board provides Cardiac
Output signal in analog form which can be displayed
on standard Oscilloscope. It will form front end module
to Patient Monitoring systems useful for the monitoring
of stroke volume and cardiac output in patients. It
is based on the principle of electrical impedance plethysmography
and is non-invasive in nature.
This method has been reported to be
accurate in majority of subjects except those with intracardiac
shunts or valvular regurgitation and at present is the
only non-invasive method for continuous cardiac output
monitoring.
Specifications
Principle
:
Electrical impedance measurement by
constant current method.
Mode of operation
:
Calibration/Patient
Configuration
:
Tetrapolar
Sensor
:
Spot electrodes
Carrier current
:
Sinusoidal at 48 KHz frequency and
4 mA amplitude
Measurement range
:
Impedance 5 to 75 W
Rate of change of impedance
:
-1.5 to +3.5 W /sec
Stroke volume
:
10 ml to 150 ml
Cardiac output
:
0.5 litres to 20 litres
Patient isolation
:
3000 Volts
Power input
:
230 VAC +/- 10%, 50 Hz, 100 W
Output
:
On Oscilloscope
Applications
This can be used for OEM application
for continuous monitoring of stroke volume and cardiac
output in critically ill patients in Intensive Care
Units and Intensive Cardiac Care Units, with hardware
addition.